Techinal Table

RADWAG MA 50.X7.IC.A.WH Moisture Analyzer — Halogen Heating to 250 °C with Automated Chamber Control

Some sample matrices only release bound moisture at temperatures well above standard ranges. The RADWAG MA 50.X7.IC.A.WH Moisture Analyzer pairs a halogen heating lamp reaching 250 °C with a 50 g capacity and fully automated drying chamber, making it a versatile tool for thermally resistant materials that require aggressive drying conditions.

Key Features

  • Halogen Heater to 250 °C: Rapid-response halogen lamp reaches high temperatures quickly, allowing efficient analysis of materials that retain moisture at lower heat levels, such as plastics and composites.
  • 50 g Sample Capacity: Sized for routine laboratory portions, accommodating a range of powdered, granular, and semi-solid materials at 1 mg mass readability.
  • Touch-Free IR Sensor Operation: Dual programmable proximity sensors open and close the drying chamber hands-free, maintaining cleanliness and preventing burns when working at elevated temperatures.
  • Automatic Motorized Chamber: Consistent opening and closing force protects delicate samples and ensures thermal stability between consecutive measurements.
  • 7-Inch Colour Touchscreen: Graphic interface shows real-time drying curves, elapsed time, and current temperature, giving operators visual confirmation of measurement progress.
  • Four Drying Profiles: Standard, fast, step, and mild modes match the heating program to each sample's thermal characteristics, preventing degradation of sensitive specimens.

Detailed Overview

The RADWAG MA 50.X7.IC.A.WH Moisture Analyzer achieves 0.001% moisture content readability, detecting subtle moisture differences critical in process optimization. The 450 W halogen module brings the chamber to 250 °C rapidly, reducing total analysis time for thermally resistant samples. Four finish modes — automatic, manual, time-defined, and user-defined — give analysts precise control over endpoint determination, whether they prefer weight-change-based automation or fixed-duration protocols. Sample traceability captures operator identity and method details, creating a documentation chain that satisfies GLP and ISO recordkeeping requirements. Connectivity through RS232, USB-A, USB-B, Ethernet, and Wi-Fi supports data export to networked systems or standalone printers for immediate hard-copy results.

Applications

  • Plastics and polymer moisture determination before extrusion or molding
  • High-temperature drying studies on composites and ceramics
  • Pharmaceutical excipient analysis requiring elevated temperatures
  • Chemical intermediate testing for tightly bound moisture

Why Choose the RADWAG MA 50.X7.IC.A.WH?

When standard 160 °C analyzers cannot fully drive off moisture from your samples, this halogen-equipped model reaches 250 °C to deliver complete, accurate results. Touch-free automation and traceable reporting make it suitable for regulated and production environments alike. Add to cart to handle your most demanding moisture analysis challenges.

BrandRADWAG
ModelMA 50.X7.IC.A.WH
Maximum Capacity50 g
Readability1 mg
Heater TypeHalogen
Temperature RangeMax 250 °C
Heating Profiles4 drying profiles (standard, fast, step, mild)
Moisture Content Readability0.001%
Moisture Content Repeatability±0.05% (sample ~2 g), ±0.01% (sample ~10 g)
Pan Sizeø90 mm, h = 8 mm
Maximum Sample Height20 mm
Display7″ graphic colour touchscreen
Weighing ChamberAutomatic
CalibrationExternal
ConnectivityRS232, USB-A, USB-B, Ethernet, Wi-Fi
Leveling SystemManual
Protection ClassIP 43
Power Supply100–120 V AC 50/60 Hz or 200–240 V AC 50/60 Hz
Heating Module Power450 W
Operating Temperature+10 to +40 °C
Net Weight5.2 kg
Recently browsed