RADWAG MA 50.R.WH MA R.WH Moisture Analyzer, 50g Capacity, 1mg Readability, Halogen, 250°C Max
RADWAG MA 50.R.WH Moisture Analyzer — Compact 50 g Halogen Analyser for Rapid Moisture Testing
Small-sample moisture determination calls for a responsive analyser that reaches target temperature quickly and resolves tiny moisture differences. The RADWAG MA 50.R.WH Moisture Analyzer handles samples up to 50 g with 1 mg readability and a halogen heater rated to 250 °C, delivering fast, reliable moisture content readings in a compact, bench-friendly footprint.
Key Features
- 50 g Sample Capacity: Sized for focused moisture work on smaller samples, minimising material usage during routine QC checks and method development.
- Halogen Heater to 250 °C: Rapid thermal ramp-up and uniform heat distribution shorten drying times, increasing daily sample throughput compared to conventional oven methods.
- 0.001 % Moisture Resolution: High-resolution moisture readings reveal subtle differences between batches, supporting tight process-control specifications.
- Informative LCD Display: A backlit screen with pictogram indicators shows active mode, connection status, and measurement progress clearly, even in variable lighting conditions.
- Multiple Connectivity Options: RS 232, USB-A, USB-B, and Wi-Fi interfaces facilitate output to printers, data-logging software, and laboratory network systems.
Detailed Overview
The MA 50.R.WH is RADWAG's lower-capacity entry in the R-series moisture analyser line, purpose-built for applications where sample sizes are small and precision matters. Moisture content repeatability of ±0.05 % with a 2 g sample and ±0.01 % with a 10 g sample ensures consistency across repeated tests. The 90 mm pan with 8 mm depth accepts standard disposable pans for quick changeovers, and the 20 mm maximum sample height covers most spreading requirements. Four drying and weighing modes — including dry-mass determination — provide flexibility for different materials and test standards. External calibration using traceable weights maintains weighing accuracy over the instrument's lifetime.
Applications
- Plastics pellet and resin moisture verification before processing
- Pharmaceutical excipient and powder moisture checks
- Cosmetics raw-material moisture screening
- Paper and packaging material moisture assessment
Why Choose the RADWAG MA 50.R.WH?
The MA 50.R.WH Moisture Analyzer is ideal for QC labs and production teams that test smaller samples frequently and need fast, dependable moisture data. Its halogen heater, wireless connectivity, and compact size make it easy to fit into any workspace. Order from Weighcore and add rapid moisture analysis to your quality toolkit.
| Maximum capacity | 50 g |
|---|---|
| Readability | 1 mg |
| Moisture content readability | 0.001% |
| Heating module | halogen |
| Heating module power | 450 W |
| Drying temperature range | max 250 °C |
| Drying temperature control | One- or two-point, depending on the drying profile (equipment: GT105k-12/Z Control Thermometer) |
| Drying temperature adjustment | Three-point adjustment at defined temperatures, programmable values (equipment: GT105k-12/Z Control Thermometer) |
| Display | LCD (backlit) |
| Display mode | %MC, %DC, %ATRO MC, %ATRO DC, g, g/kg MC, g/kg DC, -%MC |
| Communication interface | RS232¹, USB-A, USB-B, Wi-Fi® |
| Weighing pan dimensions | ø90, h= 8 mm |
| Net weight | 4.8 kg |
| Tare range | -50 g |
| Calibration | external |
| Maximum sample weight | 50 g |
| Moisture content repeatability | +/-0.05% (sample ~ 2g), +/-0.01% (sample ~ 10g) |
| Leveling system | manual |
| Maximum sample height | 20 mm |
| Protection class | IP 43 |
| Drying method | 4 drying profiles (standard, fast, step, mild) |
| Finish mode | 4 modes (automatic, manual, time-defined, user-defined) |
| Additional functions | sample traceability |
| Power supply | 100V – 120V AC 50/60Hz or 200V-240V AC 50/60Hz |
| Power consumption | 4 W |
| Operating temperature | +10 — +40 °C |
| Mass indication control | External mass standard 50 g, 100 g, 200 g (depending on the type of moisture analyzer) |
| Internal adjustment | For .IC models only, GLP report |
| External adjustment | For all models, GLP report |
| Moisture content indication control | Certified MV Test Set |
| Internal / External adjustment | Depending on the type of moisture analyzer, GLP report |