RADWAG MA 210.5Y.IC.WH MA 5Y.IC.WH Moisture Analyzer, 210g Capacity, 1mg Readability, Internal Calibration
RADWAG MA 210.5Y.IC.WH Moisture Analyzer — Internally Calibrated 210 g Halogen Analyser
Maintaining weighing accuracy over time is fundamental to trustworthy moisture results, and the RADWAG MA 210.5Y.IC.WH Moisture Analyzer addresses this with automatic internal calibration — a feature uncommon in its class. Offering 210 g capacity, 1 mg readability, a halogen heater to 250 °C, and four distinct drying modes, this instrument ensures both the drying and weighing stages of your analysis are consistently reliable.
Key Features
- Automatic Internal Calibration: Built-in calibration weights adjust the weighing system in response to temperature changes, eliminating the need for external calibration standards and reducing maintenance tasks.
- 210 g Maximum Capacity: Accommodates a full range of sample sizes and materials, from fine powders to coarser granulates, without requiring sub-sampling.
- Four Drying Modes (Standard, Mild, Fast, Step): Tailor the heating profile to each material type — rapid ramp for quick production checks, gentle curves for heat-sensitive specimens, or multi-step programmes for complex matrices.
- Automatic Drying Chamber: Motorised operation applies uniform force every cycle, improving measurement repeatability and extending the instrument's mechanical lifespan.
- 10-Inch Touchscreen Display: Drying curves, sample parameters, and instrument diagnostics are all accessible from a single, responsive interface designed for intuitive navigation.
Detailed Overview
The MA 210.5Y.IC.WH shares the 5Y platform with RADWAG's premium moisture analyser line but distinguishes itself with internal calibration, ensuring the gravimetric foundation of every moisture test remains verified without operator intervention. The halogen module delivers uniform heat across the 90 mm pan, supporting moisture content repeatability of ±0.05 % (2 g sample) and ±0.01 % (10 g sample). Semi-automatic LevelSENSING technology monitors instrument alignment, alerting operators when corrective action is needed. Connectivity options include USB, Ethernet, Wi-Fi, and hotspot for direct integration with LIMS and quality-management systems. The combination of internal calibration and an automatic chamber positions this analyser as a low-maintenance, high-confidence solution for busy laboratories.
Applications
- Pharmaceutical excipient and API moisture verification
- Food safety and shelf-life moisture compliance testing
- Plastics and composites incoming-material moisture checks
- Environmental soil and sediment moisture determination
Why Choose the RADWAG MA 210.5Y.IC.WH?
The MA 210.5Y.IC.WH Moisture Analyzer uniquely combines internal calibration with a full-featured 5Y drying platform, giving quality teams confidence that both the weighing and drying elements of their analysis are accurate and traceable. It is ideal for regulated and high-throughput environments. Order from Weighcore and bring self-calibrating moisture analysis to your lab.
| Maximum capacity | 210 g |
|---|---|
| Readability | 1 mg |
| Moisture content readability | 0.001% |
| Heating module | halogen |
| Heating module power | 450 W |
| Drying temperature range | max 250 °C |
| Drying temperature control | One- or two-point, depending on the drying profile (equipment: GT105k-12/Z Control Thermometer) |
| Drying temperature adjustment | Three-point adjustment at defined temperatures, programmable values (equipment: GT105k-12/Z Control Thermometer) |
| Display | 10″ touchscreen |
| Display mode | %MC, %DC, %ATRO MC, %ATRO DC, g, g/kg MC, g/kg DC, -%MC |
| Communication interface | USB-A ×2, USB-C, HDMI, Ethernet, Wi-Fi®, Hotspot |
| Weighing pan dimensions | ø90, h= 8 mm |
| Net weight | 5.4 kg |
| Tare range | -210 g |
| Calibration | internal (automatic) |
| Maximum sample weight | 210 g |
| Moisture content repeatability | +/-0.05% (sample ~ 2g), +/-0.01% (sample ~ 10g) |
| Leveling system | semi-automatic - LevelSENSING |
| Weighing chamber | automatic |
| Maximum sample height | 20 mm |
| Protection class | IP 43 |
| Drying method | 4 drying profiles (standard, fast, step, mild) |
| Finish mode | 4 modes (automatic, manual, time-defined, user-defined) |
| Additional functions | sample traceability |
| Power supply | 100V – 120V AC 50/60Hz or 200V-240V AC 50/60Hz |
| Power consumption | 6 W |
| Operating temperature | +10 — +40 °C |
| Mass indication control | External mass standard 50 g, 100 g, 200 g (depending on the type of moisture analyzer) |
| Internal adjustment | For .IC models only, GLP report |
| External adjustment | For all models, GLP report |
| Moisture content indication control | Certified MV Test Set |
| Internal / External adjustment | Depending on the type of moisture analyzer, GLP report |