RADWAG MA 110.R.NS MA R.NS Moisture Analyzer, 110g Capacity, 1mg Readability, Metal Heater, 160°C Max
RADWAG MA 110.R.NS Moisture Analyzer — Reliable 110 g Metal-Heater Moisture Determination
Accurate moisture content analysis starts with dependable heating and precise mass measurement. The RADWAG MA 110.R.NS Moisture Analyzer combines a 110 g capacity with 1 mg readability and a durable metal heater reaching 160 °C, making it a practical choice for routine quality control across food, chemical, and manufacturing laboratories.
Key Features
- Metal Heater Technology: The robust metal heating element provides uniform heat distribution up to 160 °C, delivering consistent and repeatable drying results across diverse sample types.
- Four Drying Profiles: Choose from standard, fast, step, and mild drying methods to match sample characteristics, reducing analysis time while protecting heat-sensitive materials from degradation.
- Backlit LCD Display: A clear, readable screen presents moisture content, dry mass, and supplementary data such as product name and tare value, keeping operators informed throughout each test.
- Flexible Finish Modes: Four finish criteria—automatic, manual, time-defined, and user-defined—give you complete control over when a drying cycle concludes, ensuring results align with your method requirements.
- Multi-Interface Connectivity: RS232, USB-A, USB-B, and Wi-Fi interfaces enable data export to printers, PCs, and network systems for documentation and compliance purposes.
Detailed Overview
Built on the MA R.NS platform, the RADWAG MA 110.R.NS Moisture Analyzer uses loss-on-drying methodology with a 0.001% moisture content readability, ensuring fine detection of even small moisture variations. The ø90 mm disposable sample pan accommodates samples up to 20 mm in height, supporting a wide range of materials from powders and granules to pastes and slurries. Sample traceability functions record test parameters and outcomes for each measurement, simplifying audit preparation and batch tracking. External calibration options with GLP reporting allow periodic verification of both mass accuracy and drying temperature, and the unit supports certified MV Test Sets for validating moisture content indications.
Applications
- Incoming raw material inspection in food production
- Moisture testing of pharmaceutical excipients and powders
- Quality control in plastics, chemicals, and building materials
- Environmental sample analysis for soil and sediment moisture
Why Choose the RADWAG MA 110.R.NS?
The MA 110.R.NS offers a straightforward approach to moisture analysis with proven metal-heater reliability and versatile drying profiles. It suits laboratories and production environments that need dependable results without unnecessary complexity. Add to cart to streamline your moisture determination workflow with this capable analyzer.
| Maximum Capacity | 110 g |
|---|---|
| Readability | 1 mg |
| Moisture content readability | 0.001% |
| Maximum sample weight | 110 g |
| Heating module | metal heater |
| Heating module power | 450 W |
| Drying temperature range | max 160 °C |
| Drying method | 4 drying profiles (standard, fast, step, mild) |
| Moisture content repeatability | +/-0.05% (sample ~ 2g), +/-0.01% (sample ~ 10g) |
| Display | LCD (backlit) |
| Display mode | %MC, %DC, %ATRO MC, %ATRO DC, g, g/kg MC, g/kg DC, -%MC |
| Finish mode | 4 modes (automatic, manual, time-defined, user-defined) |
| Weighing pan dimensions | ø90, h= 8 mm |
| Maximum sample height | 20 mm |
| Communication interface | RS232¹, USB-A, USB-B, Wi-Fi® |
| Calibration | external |
| External adjustment | For all models, GLP report |
| Internal / External adjustment | Depending on the type of moisture analyzer, GLP report |
| Internal adjustment | For .IC models only, GLP report |
| Drying temperature control | One- or two-point, depending on the drying profile (equipment: GT105k-12/Z Control Thermometer) |
| Drying temperature adjustment | Three-point adjustment at defined temperatures, programmable values (equipment: GT105k-12/Z Control Thermometer) |
| Mass indication control | External mass standard 50 g, 100 g, 200 g (depending on the type of moisture analyzer) |
| Moisture content indication control | Certified MV Test Set |
| Additional functions | sample traceability |
| Leveling system | manual |
| Protection class | IP 43 |
| Operating temperature | +10 — +40 °C |
| Power supply | 100V – 120V AC 50/60Hz or 200V-240V AC 50/60Hz |
| Power consumption | 4 W |
| Net weight | 4.8 kg |
| Tare range | -110 g |
Recently Viewed Products
Pick up where you left off.