RADWAG MA 110.5Y.NS MA 5Y.NS Moisture Analyzer, 110g Capacity, 1mg Readability
RADWAG MA 110.5Y.NS Moisture Analyzer — Touchscreen-Driven Analysis with Metal Heater
Operators who value hands-on digital control will appreciate the RADWAG MA 110.5Y.NS Moisture Analyzer and its 10-inch touchscreen interface. Designed for 110 g capacity with 1 mg readability, this instrument uses a metal heater module and offers four distinct drying profiles — all managed through an automatic weighing chamber and semi-automatic LevelSENSING system that simplify daily operation in busy quality control settings.
Key Features
- 110 g Capacity with 1 mg Readability: A practical capacity range paired with milligram precision covers a wide variety of standard moisture analysis tasks.
- Metal Heater Module: Delivers even, stable heating up to 160 °C for consistent sample drying without the replacement costs associated with halogen lamps.
- 10-Inch Touchscreen Display: Full-colour interface provides drag-and-drop parameter setup, graphical drying curves, and configurable result formats for a modern user experience.
- Four Drying Profiles: Standard, fast, step, and mild programs let operators match drying conditions to material sensitivity and moisture content levels.
- Automatic Weighing Chamber: Motorized lid operation speeds up sample loading, reduces contamination risk, and contributes to consistent measurement conditions.
- Semi-Automatic LevelSENSING: Guides operators through proper levelling to maintain measurement integrity, eliminating guesswork during instrument setup.
Detailed Overview
The MA 110.5Y.NS records moisture content with 0.001 % readability and ±0.01 % repeatability at 10 g samples, supporting demanding specifications in regulated industries. External calibration ensures weighing traceability, and connectivity via USB-A, USB-C, HDMI, Ethernet, Wi-Fi, and Hotspot integrates the analyzer with LIMS, networked storage, and peripheral devices. Multiple display modes — including %MC, %DC, %ATRO, and g/kg — provide flexibility in reporting formats. Sample traceability and GLP report generation support documentation requirements for quality audits.
Applications
- Quality control of plastics and polymers before extrusion or moulding
- Food processing moisture checks on flours, spices, and dried goods
- Pharmaceutical intermediate testing during production
- Academic research involving moisture-sensitive materials
Why Choose this moisture analyzer?
The MA 110.5Y.NS combines the durability of a metal heater with the convenience of modern touchscreen control and automated chamber handling. It is particularly suited for quality teams that run diverse sample types throughout the day and need quick access to multiple drying programs. Add to cart and modernize your moisture analysis workflow.
| Maximum Capacity | 110 g |
|---|---|
| Readability | 1 mg |
| Moisture content readability | 0.001% |
| Heating module | metal heater |
| Heating module power | 450 W |
| Tare range | -110 g |
| Maximum sample weight | 110 g |
| Moisture content repeatability | +/-0.05% (sample ~ 2g), +/-0.01% (sample ~ 10g) |
| Drying temperature range | max 160 °C |
| Drying method | 4 drying profiles (standard, fast, step, mild) |
| Finish mode | 4 modes (automatic, manual, time-defined, user-defined) |
| Display | 10″ touchscreen |
| Display mode | %MC, %DC, %ATRO MC, %ATRO DC, g, g/kg MC, g/kg DC, -%MC |
| Weighing pan dimensions | ø90, h= 8 mm |
| Weighing chamber | automatic |
| Maximum sample height | 20 mm |
| Leveling system | semi-automatic - LevelSENSING |
| Communication interface | USB-A ×2, USB-C, HDMI, Ethernet, Wi-Fi®, Hotspot |
| Protection class | IP 43 |
| Operating temperature | +10 — +40 °C |
| Power supply | 100V – 120V AC 50/60Hz or 200V-240V AC 50/60Hz |
| Power consumption | 6 W |
| Mass indication control | External mass standard 50 g, 100 g, 200 g (depending on the type of moisture analyzer) |
| Internal adjustment | For .IC models only, GLP report |
| External adjustment | For all models, GLP report |
| Internal / External adjustment | Depending on the type of moisture analyzer, GLP report |
| Drying temperature control | One- or two-point, depending on the drying profile (equipment: GT105k-12/Z Control Thermometer) |
| Drying temperature adjustment | Three-point adjustment at defined temperatures, programmable values (equipment: GT105k-12/Z Control Thermometer) |
| Moisture content indication control | Certified MV Test Set |
| Additional functions | sample traceability |
| Net weight | 5.3 kg |
| Calibration | external |